Input Specifications |
1,2,3 and 6 points |
DC voltage: ±13.8mV, ±27.6mV, ±69mV, 200mV, ±500mV,±2V, ±5V Built-in voltage divider; ±10V, ±25V, ±50VD
Ccurrent: External installation of shunt resistor(250Ω) is applied (option)
Thermocouples: K, E, J, T, R, and B (option)
Resistance thermometer: Pt100(1997) (Measured current; 1mA)* Linear scale for thermocouple and resistance thermometer |
Single scale (standard), double scale (option) |
±0.5% of input span (except for some input under standard operating condition) Refer to the table of standard range and minimum width of scale for non-standard input |
0.3% of input span |
K,E,J,T --- ±1.0°C or less (23°C±10°C) ±2.0°C or less (0 to 50°C)
(For internal reference junction compensation, the errors above are added to the accuracy rating) |
±0.02%/°C (Converted into reference ranges) |
6 seconds/point |
Approximately 1/2,000 |
On thermocouple or resistance thermometer input, disconnection of signal can be detected
.(Up-scale and down scale burnout on all channels can be programmed.)
Burnout detection: Voltage application method (approximately 8V, 1mA) |
Thermocouple inputs, DC voltage inputs (±5V or less) --- 1kΩ(burnout disabled) or less
DC Voltage inputs (input more than ±5V) --- 100Ω or less
Resistance thermometer inputs --- per wire 10Ω or less (Same resistance for 3 wires) |
Thermocouple inputs, DC voltage inputs --- ±10V DC or less
DC voltage inputs (Voltage divider built-in) --- ±60V DC or less
Resistance thermometer --- ±6V DC or less |
30V AC |
120dB or more (50/60Hz±0.1%) |
50dB or more (50/60Hz±0.1%) |
Recording Specifications |
±0.5% of recording span |
Inkpad dotting |
Input span movement --- approximately 2 seconds |
1: red 2: blue 3: green 4: violet 5: purple 6:brown |
Fan-fold type: total width of 114mm, total length of 10m, effective chart width of 100mm |
6-speed change, 5,10,20,40,80,160mm/h (standard) |
±0.1% or less (It is based on the chart scale.) |
Indicating Specifications |
Scale plate and pointer |
Single scale or double scale (minimum scale division: 80) |
Alarm Specifications |
Pointer and alarm-point seal pasted on scale.
Alarm LED lamp lightens for alarming
(All channels OR output) |
Higher and lower-limit alarm |
Individual setting for higher and lower-limit value
(Programming percentage of input span by indicating pointer, input resolution 0.5%) |
0.4% of input span |
1a contact and 2 outputs (common)
Maximum contact capacity: 2A (resistive load), 0.5A (inductive load) |
General Specifications |
100 to 240V AC, 50/60Hz (Universal power supply) with power supply switch |
Maximum 12VA (100V AC), Maximum 20VA (240V AC) |
Reference operation condition
Ambient temperature range: 21 to 25°C Ambient humidity range: 45 to 65%RH Power voltage: 100V AC±1% Power frequency: 50/60Hz±0.5% Attitude: left/right 0°, forward tilting 0°, backward tilting 0° Warm-up time: longer than 30 minutes
Normal operation condition
Ambient temperature range: 0 to 50°C Ambient humidity range: 20 to 80%RH Power voltage: 90 to 264V AC Power frequency: 50/60Hz ±2% Attitude: left/right 0 to 10°, forward tilting 0°, backward tilting 0 to 20°
Transportation condition (at the packed condition on shipment from our factory)
Ambient temperature range: -20 to 60°C Ambient humidity range: 5 to 90%RH (No dew condensation) Vibration: 10 to 60Hz, 4.9m/ S2 (0.5G) or less Impact: 392m/S2 (40G) or less
Storage condition
Ambient temperature range: -20 to 60°C Ambient humidity range: 5 to 90%RH (No dew condensation) |
Secondary terminals and protective conductor terminals --- 20MΩ or more at 500V DC Primary terminals and protective conductor terminals --- 20MΩ or more at 500V DC Primary and secondary terminals --- 20MΩ or more at 500V DC
Note:Primary terminals --- Power (L,N), Alarm terminals (mechanical relay)
Secondary terminals --- Measurement input terminals |
Secondary terminals and protective conductor terminals --- 1 minute at 500V AC
Primary terminals and protective conductor terminals --- 1 minute at 1500V AC
Primary and secondary terminals --- 1 minute at 2300V AC Notes;
Primary terminals --- Power(L,N), Alarm terminals (mechanical relay)
Secondary terminals --- Measurement input terminals |
Door (frame) --- ABS resin, window --- glass Case --- ABS resin |
Door (frame) --- Black (equivalent to Mussel N1.5),window --- Transparent Case --- Black (equivalent to Mussel N1.5) |
Panel mounting |
Approximately 1.7kg |
Indication fluctuation 0.2% or less (conversed into reference ranges at 90 to 264V AC) |
Optional Specifications |
Alarm contact output is available
Alarm relay --- Mechanical relay 1a contact, 2 outputs (common)
Maximum contact rating: 250V AC 2A, 30V DC 2A (resistive load), 250V AC 0.5A, 30V DC 0.5A (inductive load) |
250Ω of shunt resistor is applied to measure voltage input |
Built-in voltage divider(1/1000) measures input in the range of ±5V to ±50V (input type "7" only) |
Refer to the table of standard range and programmable minimum width of scale Minimum width of scale ---
DC voltage: 10mV DC width or mor
Thermocouple: K; 250°C width or moreE,J,T; 200°C width or moreR; 800°C width or more
Resistance thermometer: 100°C width or more |
Measures input with 2 types of scales (each scale is only serial channel) |
Function for detecting disconnection for sensor with thermocouple or resistance thermometer input.
Up-scale and down scale burnout on all channels can be programmed (Input type "7" only), parallel operation is not possible |
5-speed change, 5,10,20,40,80mm/minute, hour change |
Maximum length 15.6m |